EasyManua.ls Logo

Agilent Technologies EasyEXPERT - Page 453

Agilent Technologies EasyEXPERT
610 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15 Structure
15.26 R-I: Resistor R-I characteristics (A.01.11)
[Supported Analyzer]
B1500A, 4155B, 4155C, 4156B, 4156C
[Description]
Measures the resistance vs current characteristics (R-I characteristic).
[Device Under Test]
Resistor, 2 terminals
[Device Parameters]
L: Resistor length
W: Resistor width
Temp: Temperature
[Test Parameters]
IntegTime: Integration time
Port1: SMU connected to resistor, primary sweep current output
I1Start: Sweep start current
I1Stop: Sweep stop current
I1Step: Sweep step current
V1Limit: Port1 voltage compliance
Port2: SMU connected to resistor, constant voltage output
[Extended Test Parameters]
V2: Port2 output voltage
HoldTime: Hold time
DelayTime: Delay time
[Measurement Parameters]
Port1 measurement voltage V1
[User Function]
Resistance R=V1/I1
Sheet resistance Rsheet=R*W/L
[X-Y Graph]
X axis: Port1 output current I1 (LINEAR)
Y1 axis: Port1 measurement voltage V1 (LINEAR)
Y2 axis: Resistance R (LINEAR)
[List Display]
Port1 output current I1
Port1 measurement voltage V1
Resistance R
Sheet resistance Rsheet
Agilent EasyEXPERT Application Library Reference, Edition 8
15-45

Related product manuals