17 Utility
17.6 ForcePGC: PG Output1 Continuous Output (A.01.20)
[Supported Analyzer]
B1500A, B1505A, 4155B, 4155C, 4156B, 4156C
[Description]
Sets Output1 of Agilent 81110A Pulse Generator, and triggers continuous pulse output. Pulse leading/trailing
edge transition time can be set.
Execute ResetPG to stop pulse output before the specified pulses are applied.
[Required Modules and Accessories]
Agilent 81110A Pulse Generator (2-output, PGU1 and PGU2)
GPIB cable
[Test Parameters]
Address: GPIB address of Agilent 81110A Pulse Generator
Period1: Pulse period [s]
Delay1: Pulse delay time [s]
Dcyc1: Duty cycle [%]
LeadTime1: Pulse leading edge transition time [s]
TrailTime1: Pulse trailing edge transition time [s]
Level1: Pulse High level [V]
Base1: Pulse Low level [V]
ExtImp1: Load impedance [ohm]