Maintenance
7
Simple Repairs
3
Rotate the gripper arm slowly from left to right and apply a
gentle pressure to the paper clip. The clip will engage on
an internal catch and the rotation of the arm will be
blocked.
4 Hold the paper clip in place, press the gripper release
button and rotate the gripper arm to the right.
5 The gripper arm will come off. 6 Replace the gripper arm by holding the paper clip in place,
pushing the gripper arm into the holder and rotating the
gripper arm to the left.
Next Steps:
7 On completion of this procedure: Install the front cover.
8 Turn the power to the autosampler ON.