1260 Infinity II Binary LC System - Manual and Quick Guide 55
Quick Start Guide
3
Check Out the System
Find the correct settings for the individual modules here:
• Checkout method parameter settings Binary Pump (G7112B)(Ta ble 3 on page 56)
• Checkout method parameter settings Multisampler (G7167A) or Vialsampler
(G7129A)(Ta bl e 4 on page 56)
• Checkout method parameter settings Multicolumn Thermostat (G7116A)(Tab le 5 on
page 56)
• Checkout method parameter settings Diode Array Detector HS (G7117C) or Diode Array
Detector WR (G7115A)(Tab le 6 on page 57)
• Checkout method parameter settings Variable Wavelength Detector (G7114A)(Tabl e 7
on page 57)
Table 2 Overview of column options for different pumps
Pump Option 1 Option 2 Option 3
1260 Infinity II Quaternary and Binary Pump
(G7111B/G7112B)
Poroshell 120 EC-C18,
4.6 x 100 mm, 2.7 µm
Poroshell 120 EC-C18,
3.0 x 150, 2.7 µm
Poroshell 120 EC-C18,
3.0 x 50 mm, 2.7 µm
695975-902T 693975-302T 699975-302T
1260 Infinity II Quaternary Pump VL (G7111A) Poroshell 120 EC-C18,
4.6 x 100 mm, 4 µm
Poroshell 120 EC-C18,
4.6 x 50 mm, 2.7 µm
Poroshell 120 EC-C18,
4.6 x 150 mm, 4 µm
695970-902T 699975-902T 693970-902T