EasyManuals Logo

Agilent Technologies G7166A Installation Manual

Agilent Technologies G7166A
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
7
15 Connect the finger-tight fittings of the inner tubing kit to the ports of the
diverter valve. The tubings must run into the ports of the diverter valve in
horizontal, parallel lines.
NOTE
The tubings must not be bent or twisted.
16 Connect the tubing with the label with the arrow (1) to the flow cell of the
detector. Insert the other tubing (2) into the waste container.
17 Re-install the tray(s) in the tray base.
18 Start the instrument.
19 Close the front cover.



Other manuals for Agilent Technologies G7166A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7166A and is the answer not in the manual?

Agilent Technologies G7166A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7166A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals