4Service
68 U2000 Series Operating and Service Guide
Reassembly Procedure
Tools required for reassembly:
Tabl e 4- 13 Attenuator Reassembly Procedure
Tools Purpose Torq ue value
Loctite
Threadlocker 242
To secure the connection
between the attenuator and the
sensor’s connector
N/A
¾” torque wrench To tighten the attenuator 12 lbs.in
1 Apply one drop of loctite on the
first, second, and third threads of
the connector.
2 Tighten the connector using the
torque wrench.