Hook-Up Instructions
For Oxygen Concentrator to Hyperbaric Chamber
Parts:
“Christmas Tree” Connector
Connector
Connecting tubing w/Valve Nut
Step 1. Screw on the threaded end of the “Christmas Tree” connector to
the oxygen port on the Oxygen concentrator. Make finger tight.
Step 2. Push the longer end of the connecting hose on to the “Christmas
Tree” to secure.
Step 3. Remove the Brass nut from the “Head” of the Hyperbaric
Chamber. Nut is identified by a green circle around it.
Step 4. Remove the end from the shorter end of the connecting hose,
insert through the opening where the nut was removed and thread the
nut on the tubing. Tighten snugly with a wrench.
Step 5. Reach inside the chamber and pull the connecting hose through
to seat the black tape in the back side of the valve nut.
Step 6. Replace the end on to the tubing with the larger end facing out.
Twist on gently to insure there is no kinking in the tubing. The plastic
connector fits in this end and may be used to connect mask or cannula
for in chamber use.
You are now ready to use the Concentrator with the chamber.

Other manuals for AirSep newlife intensity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AirSep newlife intensity and is the answer not in the manual?

AirSep newlife intensity Specifications

General IconGeneral
Oxygen Concentration93% ±3%
Oxygen Purity93% ±3%
Operating Pressure9.5 PSI
Noise Level45 dB
Power Supply110-120V, 60Hz