20
Alpha-Stim
®
AID output waveform parameters
(Load resistance = 1000Ω)
Al
ha-Stim
®
Waveform
tem Parameter 0.5 Hz
1. Average pulse repetition rate
(
ulses
er second)
0.8
2. Pulse width (seconds) 0.25
0.5
0.75
1
3. Charge per pulse at 500 µA (µC) 125
250
375
500
4. Period (seconds) 10
5. Total charge in each direction in a
eriod at 500 µA (mC)
1.25
6. Dut
c
cle (%) 50
7. Net current in either direction 0
8. Output current
0 to 500 µA adjustable in
50 µA increments
Storage
Remove the batteries when storing the Alpha-Stim
®
AID for an extended time of
more than one month. Use the case to store and transport the
Alpha-Stim
®
AID. The Alpha-Stim
®
AID and its accessories should be stored within
a temperature range between -22⁰ C and 55⁰ C (-7⁰ F and 131⁰ F) and used within
a temperature range between 5⁰ C and 40⁰ C (41⁰ F and 104⁰ F), with a relative
humidity below 90%, at an atmospheric pressure between 912 to 1115 hPa.
Cleaning
Clean the Alpha-Stim
®
AID by gently wiping the surface of the case and screen
with a damp cloth when dirty. Use mild soap and water if necessary. Use of other
cleaning solutions may damage the case and screen. Never spray cleaners
directly on the case and screen. Between treatments, the Earclip Electrode Pads
(EEPS™) may be removed, and 70% isopropyl alcohol may be used on the Earclip
Electrodes. New EEPS™ should be placed on the Earclip Electrodes prior to the
next treatment.
STORAGE AND CLEANING