UM_AlsSUMPC(50_100_140)_EN_1.0_11-17.docx 8 / 14
TECHNICAL DATA ----------------------------------------------------------------------------------------------
Technical features
Generator: Electronic type, compliant with IEC 601-2-2 standards and suitable for monopolar and bipolar usesl
Working frequency: Monopolar/bipolar working frequency 440kHz +- 10%.
Generator PER: 97%.
Classification: (CE2007/47= IIB) (IEC= Class I - Type CF) (EMC= Cat. A).
Output circuit: “floating” insulated from earth at high/low frequencies and protected against the use of the defibrillator.
Supply, absorption and fuses: See the data label on the back.
Intended use: Up to ≤2000mt, in environment with pollution degree = Cat II, with a supply network with overvoltage = Cat 2.
Enclosure IPN
3
N
2,
= Protected against the ingress of solid objects ≥ 2,5 mm and against water drops when tilted up to 15°.
Foot-switch IPN
6
N
7
= Protected against both the dust and effects of a momentary immersion in water.
Cooling: By convection without fan.
Activation signals Cut (yellow light and grave sound), Coagulation ( blue light and acute sound)
Operation control: By microprocessor
Self-test system: By microprocessor with self-checks and ERROR CODES
Neutral electrode (single area and double area) control: Specific circuit with power delivery block and alarm signals.
Supply cord: 2meters, section 3x0,75 mm
2
.
Dimensions and weight: cm (WxDxH) 22x24x12 – about Kg 4.5
Environmental and atmospheric conditions for use, transport and storage
For the use: Temperature (°C) +10 ÷ +40. Humidity 30% ÷ 75%. Pressure (hPa): 700 ÷ 1060.
For the transport and storage: Temperature (°C) -40 ÷ +70. Humidity 10% ÷ 95%. Pressure (hPa): 500 ÷ 1060.
Currents and Powers
W: Power (Watts) - : Rated load –
Vpp: Open circuit peak to peak voltage - CF: Crest factor – M: Modulation - DT: Duty Cycle –
SU50-MPC SU100-MPC
Pure
80 W
500
Vpp 980 - CF 1.5 - M 0% - DT 100%
Pure
160 W
500
Vpp 1000 - CF 1.5 - M 0% - DT 100%
P
PULSED
40 W “ Vpp 1350 - CF 3 - M 50% - DT 100%
P
PULSED
80 W “ Vpp 1360 - CF 3 - M 50% - DT 100%
Blend
80 W “ Vpp 1400 - CF 2.3 - M 0%, DT 80%
Blend
140 W “ Vpp 1400 - CF 2.3 - M 0%, DT 80%
B
PULSED
1
40 W “ Vpp 1550 - CF 3.5 - M 50%, DT 80%
B
PULSED
1
70 W “ Vpp 1550 - CF 3.5 - M 50%, DT 80%
B
PULSED
2
33 W “ Vpp 1580 - CF 3.6 - M 50% - DT 80%
B
PULSED
2
38 W “ Vpp 1580 - CF 3.6 - M 50% - DT 80%
Micro
80 W “ Vpp 1530 - CF 3.4
M 0% - DT 50%
Micro
80 W “ Vpp 1530 - CF 3.4
M 0% - DT 50%
M
PULSED
80 W “ Vpp 1530 - CF 3.4
M 0% - DT 50%
M
PULSED
80 W “ Vpp 1530 - CF 3.4
M 0% - DT 50%
Ful
75 W
750
Vpp 2250 - CF 3.5 - M 0% - DT 50%
Ful
100 W
750
Vpp 2300 - CF 3.5 - M 0% - DT 50%
F
PULSED
40 W “ Vpp 2300 - CF 5 - M 50% - DT 50%
F
PULSED
40 W “ Vpp 2300 - CF 5 - M 50% - DT 50%
Bi
olar
80 W
100
Vpp 500 - CF 2.8 - M 0% - DT 100%
Bi
olar
100 W
100
Vpp 500 - CF 2.8 - M 0% - DT 100%
SU140-MPC
Pure
140 W
500
Vpp 1000 - CF 1.5 - M 0% - DT 100%
P
PULSED
80 W “ Vpp 1380 - CF 3 - M 50% - DT 100%
Blend
140 W “ Vpp 1400 - CF 2.3 - M 0%, DT 80%
B
PULSED
1
70 W “ Vpp 1550 - CF 3.5 - M 50%, DT 80%
B
PULSED
2
38 W “ Vpp 1580 - CF 3.6 - M 50% - DT 80%
Micro
90 W “ Vpp 1530 - CF 3.4
M 0% - DT 50%
M
PULSED
90 W “ Vpp 1530 - CF 3.4
M 0% - DT 50%
Ful
120 W
750
Vpp 2300 - CF 3.5 - M 0% - DT 50%
F
PULSED
60 W “ Vpp 2300 - CF 5 - M 50% - DT 50%
Bi
olar
100 W
100
Vpp 500 - CF 2.8 - M 0% - DT 100%