Sequoia / Tahoe Series User Manual California Instruments
M447352-01 REV-A 293
incurred, compared to the specifications of the parameters that are to be calibrated. To set up the
alignment procedures, perform the following initial steps:
1. Disconnect AC mains power when making setup connections.
2. Connect the test equipment and control inputs of the Sequoia Unit.
3. Connect the voltage source Input to the output of the Sequoia Unit.
4. Allow a 30-minute warm-up period for the power source and test equipment before conducting
the calibration procedure.