Check the centrifuge for any damages and contact the distributor if issues are found.
Position the centrifuge on a stable, level, and well-ventilated surface, at least 300mm from surroundings.
Ensure the centrifuge chamber is free from any loose pieces or packing materials before use.
Tightly seal test tubes before placing them in rotor buckets. Use only manufacturer-approved buckets.
Balance loads by ensuring opposite tubes match in weight and size. Use fluid for balance if needed.
Ensure the power switch is off and timer at 0 before connecting the power cord to a suitable supply.
Turn on the centrifuge and check for smooth operation, minimal vibration, and normal sounds. Stop if abnormal.
Always disconnect the power source before undertaking any cleaning or maintenance procedures.
Employ only high-quality test tubes to prevent fracturing and leakage of contents into rotor buckets.
Never force tubes into buckets and ensure buckets are clean. Dispose of broken glass safely.
Clean spills immediately. Use mild solutions. Rotor/buckets can be autoclaved up to 120°C.
Store centrifuge in 0-30°C, ≤30% humidity, and clear atmosphere free of dust or corrosives.
Do not store/use in unspecified conditions, with unqualified persons, or near explosive materials.
Do not overload, use damaged parts, or use non-manufacturer-approved rotors/accessories.
Avoid using chemicals causing corrosion or centrifuging flammable/explosive mediums.