OM024-6 Model 240 Operating Manual Page 11
3.6 Saving audiograms in internal memory
The user may store up to 12 audiograms (AC & BC data), referenced by
number, in the internal memory of the audiometer. To save the current set
of audiogram thresholds (the “retained” values described in Section 3.5)
press and hold the MENU key, press FREQUENCY repeatedly until
“Save Audiogram to 1” appears on screen. Use the SIGNAL keys to
select a location numbered 1-12, and then press the YES key. Release
the MENU key once confirmation appears on the display.
Note that the Save process will overwrite any records that exist in the
selected memory location.
3.7 Loading audiograms from internal memory
Press and hold the MENU key, press FREQUENCY repeatedly until
“Load Audiogram no 1” appears on screen. Use the SIGNAL keys to
select a location numbered 1-12, and then press the YES key. Release
the MENU key once confirmation appears on the display.
3.8 Printing audiograms
Three designated thermal printers (the Able AP1300 or the Sanibel MPT-
II) are available as options for use with the Model 240 audiometer. The
correct printer must be selected (use the MENU options described in
Section 3.4.2 to make this selection).
Connect the PRINTER socket on the audiometer (6-way RJ12) to the
printer with the supplied cable (refer to Section 2.5 of this operating
manual for printer set-up). Note that the printer cables for the Able
printer (A108) and the Sanibel MPT-II (A102) are not compatible.
Ensure the printer is fully charged, switched on, loaded with paper and
ready to print.
Load the desired audiogram as described in Section 3.7; to print the
current audiogram ignore this instruction.
Press and hold the MENU key and press the FREQUENCY key to
display “Print Audiogram”. Continue to hold the MENU key, press the
YES key and release the MENU key. On the prompt “Is printer
ready?” press the YES key again. The audiogram will then print. To
cancel the print operation press NO.