SPECIFICATION
Operating voltage
LOW Range: 30 to 1500 VAC; HIGH Range: 1500 VAC to
122 kVAC
Operating Temperature: 0°C to +52°C(32°F to +125°F);
RH < 90%
Storage Temperature:0°C to +52°C(32°F to +125°F)
battery removed.
Environment: < 2000m, outdoor operation
Power: 9 volt alkaline battery
Power Consumption:
OFF: 340uA
Low Detection mode: 39 mA
High Detection mode:39 mA
Duty Cycle: Continuous
Response Time: Instantaneous
Dimensions: 345(L) x 75(W) x 47(H)mm (13.5 x 3.0 x 1.8 in)
Weight: 243g (0.5 lb.)
Optional Accessory: TIC 410A Hot Stick
. EMC: EN 61326-1:2006 This product complies with
requirements of the following European Community Directives:
2004/108/EC (Electromagnetic Compatibility) and 2006/95/EC
(Low Voltage) as amended by 93/68/EEC (CE Marking). However,
electrical noise or intense electromagnetic fields in the vicinity of
the equipment may disturb the measurement circuit. Measuring
instruments will also respond to unwanted signals that may be
present within the measurement circuit. Users should exercise
care and take appropriate precautions to avoid misleading
results when making measurements in the presence of electronic
interference.