Troubleshooting
Technical Service Manual
Copyright 2009, AMS, Proprietary Information, All Rights Reserved 8-2
P/N 0010-0667, Rev. A
Message Corrective Action
Vapor footswitch disabled when in Standby Press the Ready/Standby button on the
screen or on the footswitch.
Coag footswitch disabled when in Standby Press the Ready/Standby button on the
screen or on the footswitch.
Fiber Port Overheated, try another fiber.
If problem persists, contact Customer Care Center.
Wait for device port to cool off. If problem
repeats, replace the fiber. Contact the
Customer Care Center if problem persists.
Check Fiber Card Insertion
Remove and then re-insert card
Remove and reinsert fiber card. If problem
persists, contact the Customer Care Center.
Invalid Fiber Card
10-2079 not valid for this system
The fiber attached is not a valid GreenLight
device. The part number of the invalid fiber
is displayed. Replace the fiber with one
approved for use with GreenLight XPS.
Card Read Error
Remove and then re-insert card
Remove and reinsert fiber card. If problem
persists, contact the Customer Care Center.
Fiber type does not match card type
Try another fiber or card
Fiber ID does not match Fiber ID on smart
card. Make sure that the smart card being
used is the one that was packaged with the
fiber.
Fiber Connection Error
Check fiber connection or try another fiber
Usually this is caused by a bad fiber
connector. If a second fiber gives the same
message, the fiber coupler or its cabling
may be damaged. If problem persists,
contact the Customer Care Center.
Remote Interlock Opened If the remote interlock has been connected
to the operating room door, make sure the
door is closed. Otherwise, make sure the
remote interlock bypass connector is
inserted into the socket on the rear panel of
the system.
Excessive FiberLife Activity
Please clean and inspect fiber tip
FiberLife can detect if the fiber tip is
overheating. Usually this is due to a dirty or
damaged fiber. If the problem persists after
cleaning or if the fiber is damaged, replace