8
Fluid recommendations for cold temps:
Temperature ranges 1°C -20°C: Ethanol or water/glycol mixture 1:1
Temperature ranges -20°C to -40°C: Dynalene
Fluid recommendations for hot temps:
Temperature ranges 1°C to 80C: Soft/decalcified water or distilled water only. Tap water quality is highly
variable based on local conditions.
Temperature ranges 80 °C to 200 °C: Mineral/silicone oils
Warning:
• Hard water leads to calcification in the bath.
• Ferrous water can cause corrosion.
• Chloric water can cause pitting corrosion.
• Lab distilled and deionized water can cause corrosion if used long term, even in stainless steel.
Fire or other dangers when using bath fluids that are not recommended:
Please contact Anova before using other than recommended bath fluids. Anova assumes no liability for
damage caused by the selection of an unsuitable bath liquid.
Unsuitable bath fluids are fluids which:
• are highly viscous
• have creep characteristics
• have corrosive characteristics
Filling
Take care that no liquid enters the interior of the circulator.
Recommended maximum filling level of 1.5 Inches below the rim.
Pay attention to overflow of liquid.
Do not drain the bath fluid while it is hot!