Loading...
1-7
Chapter 1 General Information
Loopback Modes: Self loopback
Level Measurements: Vp-p (± 5%), dBdsx
Data Log: Continuous, up to 48 hrs
FCN4760 Frequency Converter (S332D only)
Frequency Range 4.7 GHz to 6.0 GHz
Frequency Resolution* 10 Hz
Frequency Reference Aging: ± 1 ppm/yr
Accuracy: ± 2 ppm
SSB Phase Noise £–65 dBc/Hz
(6 GHz) @ 30 kHz Offset
Spurious Responses £–45 dBc
Input Related
Spurious Residual £–90 dBm
Responses*
Measurement Range –40 dBm to –100 dBm
Sensitivity* (displayed £–100 dBm
average noise level)
Accuracy ± 1.75 dBm typ. (± 2.25 dBm max.)*
Maximum Input Level –5 dBm
without damage
Input and Output Ports:
RF In Type N, female, 50W
RF Out Type N, male, 50W
Electromagnetic Meets European community requirements for CE marking
Compatibility
Safety Conforms to EN 61010-1 for Class 1 portable equipment
Temperature:
Operating –10°C to 50°C, 85% humidity or less
Storage –50°C to +80°C
Power Dissipation 850 mW maximum
Dimensions
WxHxD 6.6cmx10.9 cm x 3.3 cm
(2.6 in x 4.3 in x 1.3 in)
Weight <0.45 kg (< 1 lb.)
* Specifications apply when connected to the Site Master.
Table 1-1. Performance Specifications (4 of 5)