2401T-F Conductivity/Temp. Electrode
User Manual
Brief Introduction
Apera Instruments 2401T-F Conductivity Electrode is featured with a proprietary platinum black sensor
for accurate readings in a wide conductivity range up to 200 mS/cm, along with a built-in temperature
sensor for automatic temperature compensation.
Technical Specifications
Range
0 to 200 mS/cm, 32 to 122°F (0 to 50°C)
Housing
Lead-free Glass
Constant
K=1.0
Connector
BNC/RCA
Cable length
3 Ft
Dimension
ø12*145 mm
Temperature sensor
30KΩ thermistor
Operating temperature
32 to 176°F (0 - 80°C)
How to Install the Electrode
1. Find the conductivity BNC socket (where it shows Cond) on the conductivity meter; Open the
rubber cap; Insert the BNC connector of the electrode to the BNC socket while twisting clockwise
until it’s locked.
2. Find the RCA socket (where it shows temp.) on the pH meter; Open the rubber cap; Insert the
black RCA connector of the electrode to the RCA socket directly.
3. After connecting the new electrode to your meter, adjust the meter’s constant K to 1.0 according
to the instruction of your meter’s manual. Then perform a 2 to 3-point calibration to ensure the
accuracy.
How to Use the Electrode
1. There is a proper amount of deionized water in the cap on top of the electrode. The conductivity
sensor is soaked in it to keep its sensitivity.

Other manuals for Apera PC60-Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apera PC60-Z and is the answer not in the manual?