7. Troubleshooting
If your Mini-Plus Compressor Nebulizer fails to function, consult the Troubleshooting guide below. If the
problem persists, consult your equipment provider.
Problem
Cause and Solution
Device
doesn’t
operate.
• Check if plug is properly fit into an appropriate electrical outlet.
•
before using, an auto shut down may activate by built-
in thermal
protector, cool down device for 30 minutes before next usage.
•
section for details)
Weak
Nebulization
• Check for proper electrical voltage.
• Check tubing for blockage or air leakage at connection to Mini-Plus
compressor or nebulizer cup, replace as needed.
•
there is any damage, replace as needed.
• Check if filter is too dirty, replace as needed.