Revision 5 - 60 - November 2017
Preventative Maintenance Schedule
Inspect Water
Supply Lines
X
Inspect incoming lines from wall to
Rada® valve.
Cleaning &
Disinfecting System
X
Ensure Control & Selector valve is
operational.
X
Test operation for proper flow and
pressure.
X
Ensure the Receiver is properly
secured to the tub.
Tub Safety Strap
X
Inspect for wear and ensure clips work
correctly.
Thermostatic
Mixing Valve &
Thermometer
accuracy
X
Verify the mixing valve and
thermometer are operational and
providing water within acceptable
temperature ranges. Inspect for leaks
Remedy® UV
Quartz Tubes
(Whirlpool only)
X X
Remove and clean annually if using
Apollo cleaning solutions, otherwise
every six months. Verify UV indicator
lights and UV bulbs are working.
Average UV bulb life span is
approximately five years.
Air Transmitter
button (Whirlpool)
X
Ensure proper operation of the
whirlpool jets.
Inspect Door
Gaskets
X
Prolong gasket life by keeping the
door latch disengaged when bath is not
Whirlpool plumbing
& pump/motor
X
Check for water leaks.
Air Spa Blower &
air lines
X
Check for water leaks.
Tub Drain Valve &
Rod Assembly
X
Inspect for leaks; lubricate operating
rod.
CABINET/ Rapid Fill™ Reservoir (if applicable)
Reservoir Fill Hose
X
Inspect for leaks and verify
connections are tight.
Reservoir Fill Valve
& Rod Assembly
X
Inspect for leaks; lubricate operating
rod.
X
Verify accuracy.
X
Verify water shuts off at the correct
level.