SYST:ENER:DOS:TYP {<type>}
Setting of Dosage Type. Parameter is 0 for time or 1 for energy.
SYST:ENER:DOS:TYP?
Request for state of Dosage Type. Answer is 0 for time or 1 for energy.
SYST:ENER:DOS:TIM {<time>}
Setting of time (in seconds) for time dosage.
SYST:ENER:DOS:TIM?
Request for time for “time dosage”. Answer is time in seconds.
SYST:ENER:DOS:ENER {<energy>}
Setting of energy (in Ws or VAs) for energy dosage.
SYST:ENER:DOS:ENER?
Request for energy (in Ws or VAs) for energy dosage.
SYST:ENER:DOS:ENER:K {<energy>}
Setting of energy (in kWh or kVAh) for energy dosage.
SYST:ENER:DOS:ENER:K?
Request for energy (in kWh or kVAh) for energy dosage.
SYST:ENER:DOS:ENER:TYP {< type >}
Setting of energy type for Energy Dosage (Test).
Parameters are 0 = Active, 1 = Reactive, 2 = Apparent.
SYST:ENER:DOS:ENER:TYP?
Request for energy type for Energy Dosage (Test).
Answers are 0 = Active, 1 = Reactive, 2 = Apparent.
SYST:ENER:DOS:RES:SAV
Saving the results of Dial (Dosage) test. Return values are: 0 result available, 1 saved result, 2
no result available.
SYST:ENER:DOS:RES?
Request for overall error of Dial (Dosage) test after finishing this test.
SYST:ENER:DOS:DUT:ENER:K {<x>,<y>}
Setting the values of Dial Test from el. meter. Parameters are: x = value type, 0 Initial, 1 Final, 2
their difference. y = actual value of chosen type by x.
SYST:ENER:DOS:DUT:ENER:K?
Request for all values of meter dial: Initial, Final, and differential (delta).
SYST:ENER:DOS:MAXD:WIN {<window>}
Setting the value of Maximum Demand Test’s window parameter. This parameters represents setting
of Sliding or Fixed type of window.
SYST:ENER:DOS:MAXD:PER {<period>}
Setting the value of Maximum Demand Test’s period parameter. This parameters represents time
stated in minutes.

Other manuals for Applied Precision WS 2320

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Precision WS 2320 and is the answer not in the manual?

Applied Precision WS 2320 Specifications

General IconGeneral
BrandApplied Precision
ModelWS 2320
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals