6. Cleaning at home
After each use, wash the pumpset with dishwashing liquid and rinse well. Once a day, disassemble the
pumpset and boil the parts for 10 minutes or clean in the dishwasher, microwave (in the microwave bag)
or steam steriliser (follow the manufacturer’s instructions), then dry with a clean cloth and cover.
ATTENTION: After boiling, allow the parts to cool down before you touch them.
The illustrated parts can be wiped down with a cloth as needed. It is not
necessary to flush the tube.
Do not boil the bottle holder.
Do not clean the lip valve with a brush.
7. Cleaning in hospitals (only for the Clinic Pumpset)
Sterilise the disassembled pumpset at 121°C (250°F) for 20 min / 134°C (273°F) for 5 min.
8. Use
Wash your hands thoroughly. Sit in a comfortable and
relaxed position.
8.1 Expressing with one pumpset
Connect the pumpset to the Ardo breastpump.
Centre your nipple so that it moves freely in the breast
shell (breast shell insert) when pumping.
Switch the breastpump on.
8.2 Expressing with a double pumpset
Connect the second pumpset to the Ardo breastpump.
Procedure for the Ardo Carum, Ardo Alyssa and
Ardo Bellis breastpumps: Place a pumpset on the
breast, switch the pump on and place the second
pumpset on the other breast. Alternatively, you can also
attach both pumpsets at the same time.
Procedure for the Ardo Calypso breastpump: Attach
both pumpsets at the same time (see illustration). Switch
the pump on and hold each pumpset separately in your
hands.