93
Emission and Immunity Tests
The Bellis Pump is classified for following tests and test levels:
CISPR 11 group 1 class B.
IEC61000-3-2 Harmonic current emissions.
IEC61000-3-3 Voltage changes, fluctuations and flicker compliant for the full range of supply voltages from
100V-240V at 50/60Hz.
IEC61000-4-2 Electrostatic Discharge compliant for
- ± 8 kV contact discharge
- ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air discharge
IEC61000-4-3 radiated RF electromagnetic field compliant with
- 10 V/m; 80 MHz – 2,7 GHz; 80 % AM at 1 modulation
IEC61000-4-3 Immunity to proximity field from RF wireless communications equipment compliant
- TETRA 400 385MHz @ 27V/m
- GMRS/FRS460 450MHz @ 28V/m
- LTE Band 13,17 710,745,780MHz @ 9V/m
- GSM800/900, TETRA 800, iDEN 820, CDMA850 LTE Band 5 810,870,930MHz @ 28/7m
- GSM1800, CDMA1900, GSM 1900, DECT; LTE Band 1, 3, 4, 25 UMTS 1720, 1845, 1970MHz @ 28V/m
- Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7 2450 MHz @ 28V/m
- WLAN 802.11 a/n 5240, 550, 5785 MHz @ 9V/m
IEC61000-4-4 Immunity to fast transient/burst on a.c. mains
- ± 2 kV 100 kHz repetition frequency
IEC61000-4-4 Immunity to fast transient/burst on SIP/SOP
- ± 1 kV 100 kHz repetition frequency
IEC61000-4-5 Immunity to Surge
- line to line mains ± 0,5 kV, ± 1 kV
- line to ground SIP/SOP ± 2 kV
- line to ground mains ± 0,5 kV, ± 1 kV, ± 2 kV