EasyManua.ls Logo

Ardo bellis - Page 98

Ardo bellis
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
94
IEC61000-4-6 Immunity to conducted disturbances induced by RF fields on a.c. mains
- 3 V; 0,15 MHz – 80 MHz; 6 V; ISM and amateur radio bands 0,15 MHz to 80 MHz
- 80 % AM at 1 kHz
IEC61000-4-6 Immunity to conducted disturbances induced by RF fields on SIP/SOPs.
- 3 V; 0,15 MHz – 80 MHz; 6 V; ISM and amateur radio bands 0,15 MHz to 80 MHz 80 % AM modulation
IEC61000-4-8 Immunity to power frequency magnetic fields
- 30 A/m 50 or 60 Hz
IEC61000-4-11 Voltage dips, short interruptions and voltage variations immunity
- 0 % UT; 0,5 cycle @ 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°, compliant
- 0 % UT; 1 cycle and 70 % UT; 25/30 cycles Single phase: @ 0°, compliant
- 0 % UT; 250/300 cycle, charging may be temporarily interrupted.
No test deviations to the basic standard. No special points to be observed for compliance with EMC.
Device is not tested against IEC60601-2-2 ANNEX BB Equipment used in the proximity of HF surgery devices.
Not relevant for the application.
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) this
device may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.

Related product manuals