Disinfectants
DISINFECTANT
HYBRID MATTRESS
SYSTEM
TOP COVER
BOTTOM
COVER
Alcohol solution ≤ 70%
●●●
Chlorine solution ≤ 1% (10,000 ppm)
1
●●●
Quaternary ammonium solution 1920ppm
●●●
Quaternary ammonium solution, 3-15%
●●
Hydrogen peroxide solution, 3-10%.
●
1
Chlorine concentrations may vary from 250 ppm to 10,000 ppm depending on local policy and contamination
status.
NOTE
Rinse and dry thoroughly with clean water to remove residual chemicals after disinfecting with each
chemical before storage
Accessories needed for cleaning/ disinfection
• Protective glasses
• Protective gloves
• Spray bottle with cleaning solution
• Spray bottle with disinfectant solution
• Spray bottle with water
• Cloths
WARNING
• To avoid eye and skin damage, always use protective glasses and protective gloves. If contact
occurs, rinse with plenty of water. If eyes or skin become irritated, contact a physician.
Always read the material safety data sheet of the disinfectant.
• To avoid electrical shock, always disconnect the pump from the power source before cleaning and
inspecting.
• To prevent cross-contamination, always follow the disinfection instructions in this IFU.
• To avoid eye or skin irritation, never disinfect in the presence of a patient.
08.AAV.02EN_02 • 10/2021 • Page 1/3