14
Preventive maintenance schedule
Equipment operators and Service/Maintenance Personnel should hold this Preventive Maintenance Schedule. If
you require further information or a service contract, please contact your local ArjoHuntleigh representative.
The equipment is subject to wear and tear, and the following maintenance instructions must be performed when
specified to ensure that the equipment remains within its original manufacturing specification.
ArjoHuntleigh strongly advises and warns that only ArjoHuntleigh Ltd company designated parts, which are
designed for the purpose, should be used on equipment and other appliances supplied by ArjoHuntleigh Ltd, to
avoid injuries attributable to the use of inadequate parts.
FREQUENCY
Before
every use
Weekly
Every 12
months
ArjoHuntleigh SLINGS
Examine the slings, their straps and attachment clips. If any part of the
sling or its straps is frayed, or any of the clips is damaged, the sling must
be withdrawn from use immediately and replaced with a new sling.
X
Where necessary, after patient use, carry out decontamination of the
sling in accordance with ArjoHuntleigh Ltd instructions
MAX.01520.INT. and local decontamination regulations.
X
FREQUENCY
Before
every use
Weekly
Every 12
months
ArjoHuntleigh STEDY
Carry out a visual inspection of all external parts. All functions must be
tested for correct operation, to ensure no adverse damage has occurred
during previous use.
X
Where necessary, after each patient use, carry out decontamination of the
ArjoHuntleigh Stedy in accordance with ArjoHuntleigh Operating, Prod-
uct Care, and Maintenance Instructions.
X
Make sure all casters rotate freely and the two rear brakes lock. X
Examine all exposed parts, especially where there is personal contact
with the patient’s body. Make sure no cracks or sharp edges have devel-
oped that could cause patient or user injury or have become unhygienic.
Replace where necessary and decontaminate.
X
Make sure all external fittings are secure and that all screws and nuts are
tight.
X
WARNING:
The points on this checklist are the minimum the manufacturer recommends. In some cases
more frequent inspections should be carried out. Continuing to use this equipment without
conducting regular inspections or continuing to use this equipment when a fault is found will
seriously compromise the user and resident/patient’s safety. Local regulations and standards
may be higher than the manufacturers. Preventive maintenance specified in this manual can
prevent accidents.
Preventive Maintenance Schedule