EasyManua.ls Logo

Astronics Trig-Tek 203M - Calibrator Adjustments

Default Icon
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Publication No. 980978 Rev. A 203M User Manual
Astronics Test Systems Calibration Procedure 4-3
Calibrator Adjustments
1. Place the CALIB-OPER switch to CALIB. (The CALIB light should illuminate.)
2. Connect the counter and Voltmeter to the CALIBRATOR OUTPUT BNC on the
front panel.
3. Observe a 139.9 ±0.4 Hz indication on the counter.
4. Set the 1 Volt front panel LEVEL ADJ for a 1.000 ±0.005 Volt RMS indication
on the Voltmeter. (The front panel LEVEL ADJ screw will need to be removed
if adjustment is necessary).
5. Place the sensitivity switch to 1-11, the Dial to 10.0, the FULL SCALE switch to
100, and the PEAK-RMS switch to PEAK.
6. Set the CAL ADJUST R9 for 100.0 ±0.8 g’s counts indication on the front
panel meter.
7. Set the Dial for 1.00.
8. Set the DIAL CAL ADJ R2 for a 100.0 ±0.8 g’s counts indication on the front
panel meter.
Figure 4-1, 203M Adjustment Locations
R8
R1
R2
8
DVM
Adjustment
R9
8
R4
R7
S2