Characteristic Specifications Supplemental Information
GENERAL/ENVIRONMENTAL
Power Requirements
95 to 264 Vac, 50–60 Hz, with safety
ground via approved power cord,
160 VA max
No range switching or fuse changes
required over the full operating range.
Temperature Range
Operating C to +45° C
C to +40° C for APx586 only
Storage –40° C to +75° C
Humidity
10 % to 80 %, non-condensing
Max Operating Altitude
3,000 m [9,840 feet]
Stabilization Time
20 minutes Allow up to 1 hour per 10°C if unit has
been exposed to a significant change
in temperature. Allow 24–48 hours to
recover if condensation has occurred.

Questions and Answers: