104 Apr-09
Avantes www.avantes.com info@avantes.com
Designations and Functions of Panel Controls
1) Front side: Fiber optic Connector SMA 905 compatible
LED shows ON/OFF Status of Lamp
ON/OFF-Switch
2) Backside: Power Input 12VDC, Inside +12VDC, outside GND
HgAr or Neon Lamp replacement
1. Plug out the power connector from the socket.
2. Remove screw protection caps on the frontside
3. Loosen 2 screws with philips screwdriver
4. take out the frontplate and electronics board
5. untighten 2 screws from HgAr socket
6. slightly bend the electrical wires out of lamp
socket
7. take out the HgAr/Neon Lamp from Lamp holder
8. replace by new Lamp, make sure the lamp legs
make no short circuit.
9. Slide back electronics board and front plate, tighten screws and put back screw