86 Apr-09
Avantes www.avantes.com info@avantes.com
extension will be *.fbr. The name of the intensity calibration file can be entered after clicking
10. Switch to the Irradiance Chart TAB to enter the hardware setup and select the colorimetric,
radiometric, photometric and/or peak parameters of interest. Then click OK.
11. Measure the output parameters in the experiment. If needed, change the integration time, such
that the maximum in Scope Mode is around 14000 A/D Counts . Block the light path to the
spectrometer, and save a dark spectrum. If the (ir)radiance of the light to be measured needs to
be displayed against time, click the time measurement TAB in the settings dialog as described in
section 4.4.6.
12. The intensity calibration as performed under point 9 can be loaded in future experiments by
After loading an intensity calibration, a dark spectrum needs to be saved before switching to
Irradiance mode.
Technical Data
Calibration Accuracy
Repeatability
Calibration Relative
Uncertainty to NIST standard
± 9.5 % (350-1100nm)
± 6.5% (1100-1950nm)
± 9,5% (1950-2500nm)
Ordering information
NIST traceable Halogen Lamp with CC-UV/VIS-CAL diffuser, incl. PS-
24V/1.25A
NIST traceable Halogen Lamp for use with AvaSphere-30-IRRAD, incl. PS-
24V/1.25A
NIST traceable Halogen Lamp for use with AvaSphere-50-IRRAD, incl. PS-
24V/1.25A
NIST traceable Halogen Lamp for use with AvaSphere-80-IRRAD, incl. PS-
24V/1.25A
AvaLight-HAL-CAL recalibration service 350-1095nm
AvaLight-HAL-CAL extended or recalibration service 1100-2500nm