16
Unscrew the red transport housing from the reference
electrode and check that the o-ring on the electrode is
properly seated. Save the transport housing for
storage of the reference electrode in the event the
analyzer is turned off or taken out of service for any
reason.
See illustration 2-5.
Carefully screw the reference electrode into the
reference electrode housing and place it with the other
electrodes.
See illustration 2-6.
Slide the measuring chamber forward until it locks in
the front position. Unclamp the left electrode holder
by moving the clamp forward.
See illustrations 2-7 and 2-8.
2-5. Transport housing removal
2-6. Installing reference electrode
2-7. Slide chamber forward 2-8. Unclamping electrode holder
red transport housing
reference electrode
reference electrode
housing