EasyManua.ls Logo

AVL 9180 - Page 50

AVL 9180
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
To perform daily cleaning, you will communicate
through the keypad interface. The prompt Perform
Daily Cleaning? will be displayed. Press YES to
accept.
The prompt Open Sample Door Introduce
Sample will be displayed. Pour a small amount of
AVL Cleaning Solution A into a clean sampling
container.
Lift the sample door. The prompt Introduce
Sample will be displayed and the pump will begin to
aspirate. Introduce the cleaning solution to the probe.
See illustration 2-20.
Hold the solution under the probe until the prompt
Wipe Probe Close Sample Door is displayed.
Use a lint-free tissue to remove the cleaning solution
from the probe, then close the door.
See illustration 2-21.
The analyzer will now display Thank You! and a
brief countdown will begin, indicated by the clock in
the lower right of the display. While the countdown is
running, open the bottle of AVL Electrode Condition-
ing Solution and pour a small amount into a clean
container.
After the countdown is completed, the system will
prompt Perform Daily Conditioning? Press YES.
2-20. Introducing cleaning solution
2-21. Cleaning probe

Other manuals for AVL 9180

Related product manuals