29
Note: AVL ISE-trol Electrolyte Controls are spe-
cially designed for your AVL Electrolyte Analyzer
and should be used exclusively to ensure absolute
quality and consistency of your patient samples.
To enter the lot number, proceed from the display
Program QC Level 1 Ranges? Enter YES and
the analyzer will prompt Current Lot: 0000
Change Lot#?
Note: The first time QC lot number information is
entered, the AVL Electrolyte Analyzer will display a
default lot number. Thereafter, the current lot num-
ber will appear.
Note: If you do not desire to change current lot
information, but wish to verify current programmed
QC ranges, press NO.
Press YES and Print old Values and Statistics?
will be displayed. Select YES to receive a printout
that contains information including the mean, standard
deviation (SD) and coefficient of variation (CV) of
stored data, or press NO to decline.
The analyzer will display New Lot! Delete Old
Data? Select YES to continue entering new lot
number information or press NO, if you decide to
keep the current lot number and data.