CHROME Manual
Model Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
CHROME-1KG
0.2g 500g OIML M2
IX. Adjustment
1. While the scale is on, press and hold the UNIT key
until the display shows CAL, then release.
2. Press the UNITkeyoncemore.Thediaplywillash
“CAL” followed by “500.0g”.
3. Place a 500 gram calibration weight on the center
of the platform.
4. Thedisplaywillash“PASS”followedby“500.0g”.
Calibration is complete.
X. Troubleshooting
• Problem: Scale will not turn on
• Solution: Check the batteries
• Problem: Displayedweightuctuatesrandomly
• Solution: Make sure your workspace is stable and
free of any air currents or vibration
• Problem: Displayed weight is inaccurate
• Solution: Perform a calibration adjustment
XI. Error Codes
1. OUTZ - Zero range has shifted. Perform calibration
adjustment. If the problem persists, the load cell may
be permanently damaged due to overloading.