VIII. Price Computing
1. Press the K key to select a unit price preset
2. Place the item to be weighed on the platform
3. The display will show the unit price, weight, and
total price.
4. Press MODE if you wish to change the weighing unit
IX. Calibration
Initial calibration is performed at the factory. However, the
calibration can drift slightly due to stresses from shipping.
Also, due to changes in local acceleration of gravity around
the world, adjustment is recommended when moving the
scale from one location to another. To check calibration:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart on next page), adjustment is
required.
Model Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
PC-201 0.02g 200g OIML M2
PC-2000 0.2g 2000g OIML M2
X. Adjustment
WARNING: If you do not use the correct weight
listed in the above chart for your scale, you WILL
miscalibrate and lose the factory calibration.
Calibration requires an accurate weight.
1. Press the CAL button to enter calibration mode. The
display will show “CAL”.
2. Make sure no weight is on the platform then tap