the CAL key again to set the zero point. The LCD
willash“CAL”followedbytherequiredcalibration
weight again (see chart on previous page).
3. Place the required weight on the center of the scale
platform.Thedisplaywillash“PASS”thenreturnto
normal weighing. NOTE: PASS only indicates that
calibrationhascompletedandisnotaconrmation
that you have used the correct weight.
4. Test the accuracy by placing the same calibration
weight back on the platform and checking the
value. If the value is within the acceptable tolerance
limit, calibration is complete. If not, try repeating
calibration under more stable conditions.
XI. Troubleshooting
• Problem: Scale will not turn on
• Solution: Check the batteries and battery contacts.
• Problem: Displayedweightuctuatesrandomly
• Solution: Make sure your workspace is stable
and free of any air currents, vibration, and static
electricity. Nearby radio wave emitting devices can
causeuctuations.
• Problem: Displayed weight is inaccurate
• Solution: Perform a calibration adjustment. If
you are still having trouble, contact our warranty
department.
XII. Error Codes
1. OUTZ - Zero point has drifted out of the acceptable
range. Try to perform a calibration adjustment. If the