Normal Operating Temperature Testing
After correcting for ambient temperature, the tested normal 5.
operating temperature should meet the limits specified in
Table 2.
If the HIGH temperature setting is not within specified 6.
limits, adjust the normal operating temperature. See the
Normal Operating Temperature Adjustment section.
If the HIGH temperature setting is within specified 7.
limits, verify the MED and then LOW settings against
specified limits (allow the warming unit to operate at each
temperature setting for about 3 to 5 minutes
before test).
Turn the warming unit to STANDBY. Remove the hose 8.
end from the port on the temperature test device.
Record the maintenance action taken and the hours of 9.
operation (read from the hour meter) in the Maintenance
Log section.
Table 2. Model 500 Series Operating Temperatures
(Operating temperature = Average temperature at the end of the hose.)
HIGH: 43° ± 3°C (109.4° ± 5.4°F)
MED: 38° ± 3°C (100.4° ± 5.4°F)
LOW: 32° ± 3°C (89.6° ± 5.4°F)
Note: Air temperature reaching the patient is approximately 2°C lower than
the listed temperatures.
Bair Hugger Model 505 Service Manual