Inating the Balloon Catheter
1. Fill the ination device (
eagle™ ination device) with sterile, liquid media.
2. Attach the ination device to the balloon lumen.
3. Open the stopcock and inate the balloon.
4. Once dilation has been attained, advance the sheath over the balloon.
Warning: Do not use air or any gaseous substances as a balloon ination media, always use sterile
liquid media.
Note: The use of an ination device with a pressure gauge is highly recommended to make sure
adequate pressure is applied and the maximum limits of the balloon are not exceeded.
Note: Slight reduction in balloon pressure may be required if resistance is encountered when advancing
the sheath over the balloon.
Caution: Do not exceed the recommended rated burst pressure (RBP) for this device. Balloon rupture may
occur if the RBP rating is exceeded. Please refer to the device label for the recommended RBP.
Deating the Balloon Catheter
Deate the balloon using an ination device. Since deation times vary based on balloon sizes and lengths,
check uoroscopically to conrm deation before attempting to withdraw.
1. Attach the ination device (
eagle™ ination device) and remove the solution from the balloon by pulling
back on the ination device.
2. Remove the ination device from the balloon catheter. This will allow ambient pressure to enter,
relaxing the balloon.
3. Gently withdraw the catheter. Use of a gentle counterclockwise twisting motion is recommended when
removing the catheter.
Warning: If resistance is felt when removing either the catheter or the guidewire from the working
sheath, stop and consider removing them as a single unit to prevent damage to product. Applying
excessive force to the catheter can result in tip breakage or balloon separation.
Using the Refolding Tool
The balloon catheter is supplied with a refolding tool. This aids in preparing the balloon for another insertion
during the procedure. To use:
1. Manually compress the balloon.
2. Position the refolding tool at one end.
3. Twist the refolding tool counter clockwise and push down on the balloon until it goes the entire length of
the balloon.
4. Once the balloon is folded, remove the refolding tool and proceed with the procedure.
Storage
Store in a cool, dry and dark place. Do not store near radiation or ultraviolet light sources as these may
damage product materials.
Bard, Eagle and X-Force are trademarks and/or registered trademarks of C. R. Bard, Inc.