Questions and Answers

  • J
    Jonathan CardenasAug 9, 2025
    Why is the temperature readout on my Barnstead International Furnace inaccurate?
    • J
      Jamie CobbAug 9, 2025
      If your Barnstead International Furnace displays inaccurate temperature readings, several factors could be responsible. An oxidized or contaminated thermocouple, a poor thermocouple connection, or reversed thermocouple connections can cause this. Ensure proper loading procedures and adequate ventilation around the furnace base. Static charge on the pyrometer case or control calibration issues may also lead to inaccurate readings. Replace the thermocouple, tighten connections, use correct loading procedures, clear the area around the furnace base, and reconnect the thermocouple correctly. If the issue persists, consider re-tuning the control or verifying and correcting all parameter and configuration values. If 'malfunction' continues, replace the control.
  • K
    Kyle WalshAug 14, 2025
    What to do if the element keeps burning out in my Barnstead International 1260 Furnace?
    • W
      Wayne HarrisAug 14, 2025
      Repeated element burnout in your Barnstead International Furnace can stem from overheating, a control malfunction, or using the incorrect element. An oxidized thermocouple leading to inaccurate readings or improper wiring can also be culprits. Ensure the furnace operates below its maximum temperature and closely monitor the control settings. Replace the control, install the correct element, and replace the thermocouple. Check the wiring diagram to ensure your furnace is wired correctly.
  • R
    rebeccaweissAug 19, 2025
    Why is my Barnstead International 1260 heating up so slowly?
    • J
      James EllisonAug 19, 2025
      If your Barnstead International Furnace is heating up slowly, it could be due to low line voltage, a heavy load in the chamber, the wrong heating element, or a low SPrr setting. Install a line of sufficient size and proper voltage, isolating the furnace from other electrical loads. Lighten the load in the chamber to improve heat circulation and install the correct element. Increase the SPrr setting.
  • T
    Thomas FreemanAug 25, 2025
    What to do if my Barnstead International 1260 Furnace does not heat up?
    • C
      Clayton HayesAug 25, 2025
      If your Barnstead International Furnace isn't heating, it could be due to a lack of power. Check the power source, fuses, and breakers. Alternatively, the thermocouple might be open or have reversed leads, the controller might be malfunctioning, the element could be burned out, the solid-state relay might be defective, or the chamber section might not be connected to the base. Replace the thermocouple or check its connections. Verify and correct all parameters and configuration values. Replace the element, replace the solid-state relay, and reconnect the chamber section to the base.
  • A
    Alyssa ParkerAug 29, 2025
    How to fix no display on Single Set Point Control of Barnstead International 1260?
    • P
      Peter AlexanderAug 30, 2025
      If there is no display on the Single Set Point Control of your Barnstead International Furnace, the ON and OFF power switch may be defective. Replace the power switch.
  • B
    Brian ThomasSep 3, 2025
    Why doesn't the power switch light illuminate on my Barnstead International Furnace?
    • P
      Patricia RobertsSep 3, 2025
      If the power switch light on your Barnstead International Furnace does not illuminate, the furnace may not be connected to a power supply. Check the furnace connection to the power supply.