Chapter 11: Specications
Clinician’s Operation Manual 11-7
Serial Data Output
Power Input and Data Connector
Data transferred through this connector is in a proprietary BCICP1030 format. Format available
upon request.
Sensor Connector
Data Type: ASCII comma delimited string output at 1 Hz. Lines terminated with a
carriage return.
Serial Data Protocol: RS232C, 9600 baud, 1 start bit, 8 data bits, 1 stop bit, no parity.
Power Requirements
Battery
Type:
AA (LR6) disposable:
Custom rechargeable:
4 Alkaline
Lithium-Ion (L+), 7.4V replaceable rechargeable battery pack.
Use Time:
AA (LR6) disposable:
Custom rechargeable:
WW1020 - Approximately 26 hours continuous use
WW1000 - Approximately 32 hours continuous use
WW1020 - Approximately 31 hours continuous use (new)
WW1000 - Approximately 54 hours continuous use (new)
Charge Time:
AA (LR6) disposable:
Custom rechargeable:
Not rechargeable.
Fast charges in approximately 3 hours. Charge time aected by input
power source and device loading. See Chapter 4: Operating Instructions.
Charge Cycles:
(Custom rechargeable)
300 to 80% capacity
AC Charger
WW1095 30 Watt AC Power Supply: 9V, 3A output, Input of 100-240 VAC 50Hz, 60Hz.
USB
WW1089 USB Interface Cable: Input: 5V, 500mA max (USB powered source )
Output: 9V, 230mA max to oximeter.
Length: Maximum 5 meters (16.4 feet)
Monitor Dimensions
Width: 85 mm (3.3 inches)
Height: 154 mm (6.1 inches)
Depth: 45 mm (1.7 inches)
Weight: 340 grams (12 ounces) with four (4) AA batteries
369 grams (13 ounces) with rechargeable battery pack