WARNINGS
Read the instruction manual carefully prior to setting up and using the
pressure pump. The pressure built up internally during use can be
very high.
Do not generate higher pressures than 20 bar (300 psig).
Higher pressures may damage the pump.
Only personnel with good experience and knowledge of high pressure
are allowed to work with the pressure pump. Incorrect use may result
in damage to the pump, the instrument connected to the pump and/or
personal injury.
Ensure that all connections are made correctly and that the hose and
the connectors are intact. Do not use faulty hoses or connectors.
Do not connect the pump to a high pressure source.
Use only the connectors provided with the pump. Impurities from
wrong materials may plug the pump.
Dear user,
We have made every effort to ensure the accuracy of the contents of this
manual. Should any errors be detected, we would greatly appreciate to re-
ceive suggestions to improve the quality of the contents of this manual.
The above not withstanding, we can assume no responsibility for any errors
in this manual or their eventual consequences.
We reserve rights to make modifications to this manual without any further
notice.
For more detailed technical data about the Calibration Pressure Pump,
model PGM, please contact the manufacturer.
8801300/UEPGM/000528
APPLICATION EXAMPLES
Pressure
instrument
T/C, Low V
4-w meas
R, RTD
3 & 4-w m eas
V, I,
+24V I
MEAS/SINK
V, HART
®
Low V
MEAS & IOUTPUT
GEN
SENSOR M EASURE & SIMU LATE
ET E
T/C INT. RJ
T/C OR EXT
WIRES ONLY
?
2
1
0
3
4
5
6
7
8
9
DCBA
2-w xmtr
MULTIFUNCTION CALIBRATOR
Pressure
pump
Calibrator
06.10.2000 11:49
Input
Error
1.00
Output
Gauge
bar
8.0310
0.28 % of span
0.4982
+
0
-
0% 50% 100%
Pressur e [P1: INT6C]
Current [E: Meas]
mA
Pause
Readings
Accept
Desired Input Value 0.5000
T/C, Low V
4-w meas
R, RTD
3 & 4-w meas
V, I,
+24V I
MEAS/SINK
V, HART
®
Low V
MEAS & IOUTPUT
GEN
SENSOR M EASURE & SIMU LATE
ET E
T /C IN T . RJ
T/C OR EXT
WIRES ONLY
?
2
1
0
3
4
5
6
7
8
9
DCBA
2-w xmtr
M UL T IF U N C T IO N C AL IBR AT OR
Calibrator
Pressure
pump
Pressure
instrument
06.10.2000 11:49
Inp u t
Error
1.00
Output
Gauge
bar
8.0310
0.28 % of span
0.4982
+
0
-
0% 50% 100%
Pressure [P1: INT6C]
C u r re n t [E : M e as ]
mA
Pause
Readings
Accept
D esire d Inpu t V alue 0 .50 00
(plugged)