21
86076 DB-en/03 355.00032
English
LaserStar T Plus
en
Lamp settings
Fig. 14 Lamp settings
The [ Change ] button opens the following dialog:
Lamp type
Use the [ - ] / [ + ] buttons to select the lamp type
installed in the system.
Lamp offset
Use the [ - ] / [ + ] buttons to change the lamp
offset value.
Lamp pulse counter
The [ Reset ] button resets the counter to zero (after
lamp replacement).
It is recommended to confirm the replacement
otherwise the system generates repeatedly inappro-
priate maintenance recommendations.
Lamp warn level
The [ Set ] button opens the
ALPHANUMERIC
INPUT WINDOW to enter the number of pulses
used as threshold for the maintenance recommen-
dation for a lamp replacement.
Change water filter
The [ Set ] button opens the
ALPHANUMERIC
INPUT WINDOW in order to confirm the
replacement of the water filter.
It is recommended to confirm the replacement
otherwise the system generates repeatedly inappro-
priate maintenance recommendations.
If you confirm the replacement of the water filter
without having replaced it the laser may be
damaged.
NOTE
In the course of time the energy of laser pulses is
reduced due to aging effects of the laser flash
lamps. You can compensate this effect by adding
an offset to the lamp voltage.
Furthermore you can use the lamp offset, to
calibrate the laser pulse energy when using
different laser systems or after a lamp
replacement. In this cases negative lamp offsets
may be useful.