38
Press and press simultaneouslythe tester is switched on and it enters the software
calibration state.
Preset the impact direction as .
Measure 5 points on the standard hardness block.
It will display average value after measurement.
Press to input nominal value.
Press to finish calibration.
Range of adjustment: ±15HL.
6.12 Battery replacement
Battery replacement see figure 5.
Calibration
0/5 Times
Average=550
Nominal=
550