What to do if the travel adjustment and the spreading of the carrier frame of the Beka Hospitec NORA Alu Medical Equipment do not function?
- Wwilliam56Sep 9, 2025
If the travel adjustment and the spreading of the carrier frame do not function, you should: 1. Check if the emergency stop switch is released or pressed. 2. Check that the cables of the control box are correctly plugged in. 3. Check the battery’s charging state. 4. Remove the battery and check the contacts for damage.