EasyManua.ls Logo

BENSON DT-9205A - Page 19

BENSON DT-9205A
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 18 -
AC Current
Range
Accuracy
20mA
±(1.2%rdg. + 5 digits)
200mA
±(2.0%rdg. + 5 digits)
20A
±(3.0%rdg. + 10 digits)
Overload protection: 0.2A/250V fast acting fuse. 20A range is not fused. For 20A input
terminal, each measuring time should not more than 10 seconds within interval of 15
minutes.
Indication: Average rms of sine wave
Resistance
Range
Resolution
Accuracy
200Ω
0.1Ω
±(1.0%rdg. + 8 digits)
2KΩ
0.001KΩ
±(0.8%rdg. + 5 digits)
20KΩ
0.01KΩ
±(0.8%rdg. + 5 digits)
200KΩ
0.1KΩ
±(0.8%rdg. + 5 digits)
2MΩ
0.001MΩ
±(0.8%rdg. + 5 digits)
20MΩ
0.01MΩ
±(1.2%rdg. + 20 digits)
2000MΩ
1MΩ
Reading is for reference only
Continuity test
Model
Description
All models
The beeper comes on continuously acoustic signal and the continuity
indicator LED will light for resistance <70Ω.