BestCode Next Series 8 Technical Manual October 2022 Page 197 of 290
Step 3:
Run the “Test Add
Makeup” script 3 times
by pressing the button,
and waiting for the script
to run. The repeating
until 3 full script cycles
have occurred.
Note: Jet must be on.
Step 4:
Measure the remaining volume of Makeup in the graduated
beaker
Step 5:
Determine the change in the volume of Makeup. Divide that number by 3 to determine the
volume of Makeup Used per Add Makeup routine
Results
The minimum volume per add must be greater than 1.5ml for the system to maintain it’s
viscosity. If the volume per add is less than 1.5ml, there is a fluidic problem and it should be
diagnosed using the fluidic diagrams and troubleshooting help here.