30
10. What if there are problems?
Error
message
Possible cause Solution
E1
The cuff was not
attached cor-
rectly, the cuff
line has not been
inserted properly,
or inflation takes
longer than 25
seconds.
Please wait one minute
and repeat the measu-
rement, making sure to
observe the information
in chapter “Attaching the
cuff”.
Please also check whe-
ther the cuff line has
been inserted properly
and ensure that neither
your arm nor other heavy
objects are pressing on
the line, and that the line
is not bent.
E2
The inflation pres-
sure is higher than
300 mmHg.
Please take another mea-
surement to check whe-
ther the cuff can be cor-
rectly inflated.
Make sure that neither
your arm nor other heavy
objects are pressing on
the line, and that the line
is not bent.
E3
You moved or
spoke during the
measurement in
addition to E3,
the heart rhythm
symbol is also
shown on the dis-
play.
Please wait one minute
and repeat the measu-
rement.
Ensure that you do not
speak or move during the
measurement.
E3
The blood pres-
sure values are
unusually high
or low.
The batteries are
almost empty.
Insert new batteries into
the device.
11. Disposal
Repairing and disposing of the device
•
Do not repair or adjust the device yourself. Proper operation
can no longer be guaranteed in this case.
•
Do not open the device. Failure to comply will invalidate the
warranty.
•
Repairs must only be carried out by Customer Services or
authorised retailers. Before making a claim, please check the
batteries first and replace them if necessary.