71
Test Standard Test Item Test Level Electromagnetic environment – Guidance
IEC 61000-3-2 Harmonic Current Emission Not applicable (Note 1) –
IEC 61000-3-3 Voltage Fluctuations and Flicker Not applicable (Note 1) –
Note 1: The power supply of this device is from battery (DC voltage)
A1.3 Guidance and manufacturer’s declaration – Electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure
it is used in such an environment.
Test Standard Test Item Test Level Electromagnetic environment – Guidance
IEC 61000-4-2 Electrostatic discharge Contact:
± 8 kV;
Air: ±2 kV, ±4 kV,
±8 kV, and ±15 kV
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Portable and mobile RF communications
equipment should be used no closer than 30 cm
(12 inch) to any part of this device.
IEC 61000-4-3 Radiated RF EM fields 10 V/m;
80 MHz to 2.7 GHz;
80% AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer than 30 cm
(12 inch) to any part of this device.
IEC 61000-4-3 Proximity fields from RF
wireless communications
equipment
Not applicable
(Note 1)
–
IEC 61000-4-4 Electrical fast transients /
bursts
Not applicable
(Note 1)
–
IEC 61000-4-5 Surges line-to-line Not applicable
(Note 1)
–
IEC 61000-4-6 Conducted disturbances
induced by RF fields
Not applicable
(Note 1)
–
IEC 61000-4-8 Rated power frequency
magnetic fields
30 A/m
50 Hz and 60 Hz
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.