18
www.bindicator.com
TDR180003 Rev. I
Name (number), function Adjustable value range Description
Default value
Maximum measurement range limit
(P06),
Remote blocking
0 (limiter off) or the minimum
measuring distance. The distance
between (P05) +5 cm and the probe
length (P03)
The maximum value of the
measuring range of the device
can be limited with this parameter.
Signals received beyond the set
distance will not be processed. Its
value is calculated from the plane
of the lower sealing surface of the
mechanical connector (in the case
of a flanged version, from the lower
plane of the flange).
Remote blocking can be turned off
by setting the parameter to 0. In
this case, the device will give a valid
result along the entire length of the
sensor.
MAIN MENU / MEASUREMENT
CONFIG / MAX. MEAS. DIST.
0
(off)
Damping time (P20)
Time constant
0…999 sec The “Damping time” is used to
reduce undesired fluctuations
(e.g.: rippling) in the display of the
measured data. If the level changes
abruptly, the new value will be set
to within 1% accuracy (exponential
setting). Unit of measure: seconds.
MAIN MENU / MEAS.
OPTIMIZATION / DAMPING TIME
10 sec
Probe length (P03)
0.1…30 m
(0.33…100 ft)
When the current output is set
to “Fixed” mode (P12b = 1), the
constant current set here is set.
The PV current transmission does
not work, and at the same time, it
overrides any current output fault
indication.
MAIN MENU / OUTPUT SETUP
/ ANALOG OUTPUT / MANUAL
VALUE
This distance plus an additional
0.66 ft (0.2 m). It is the hard limit of
the measurement. (Excess is used
up for probe end detection and
pollution of the probe.)
Fixed output current (P08),
Setting constant output current
3.8…22 mA When the current output is set
to “Fixed” mode (P12b = 1), the
constant current set here is set.
The PV current transmission does
not work, and at the same time, it
overrides any current output fault
indication.
MAIN MENU / OUTPUT SETUP
/ ANALOG OUTPUT / MANUAL
VALUE
4
System of measurement units, units
of measurement:
Path of parameters: “Settings” /
Application” tab