17
Does not contain hazardous substances requiring special handling and
treatment
Dispose of in an environmentally responsible manner in accordance
with regional requirements
Contact Bio Compression Systems if you have questions or concerns
regarding disassembly and disposal
Manufacturer’s EMC Declaration
Electromagnetic Emissions
Emissions
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The device uses RF energy only for its internal functions. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low voltage power supply
network that supplies building used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not
applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not
applicable
Electromagnetic Immunity
Immunity Test
Immunity Test Level
IEC 61000-4-2 Electrostatic Discharge Immunity
±8kV contact, ±2, 4, 8, 16kV air discharge
IEC 61000-4-3 Radiated RF Field Immunity
80MHz 2.7GHz, 10V/m, AM 80% at 1kHz
IEC 61000-4-3 Proximity Fields from RF Wireless
Communications Equipment
IEC 60601-1-2, Section 8.10, Table 9
IEC 61000-4-4 Electrical Fast Transients
±2kV/100kHz power, ±1kV/100kHz signal
IEC 61000-4-5 Surge Immunity
±0.5, 1kV line to line, ±0.5, 1, 2kV line to ground
IEC 61000-4-6 Conducted RF Immunity
150kHz - 80MHz, 3V
RMS
in whole range, 6V
RMS
in amateur radio and
ISM, AM 80% at 1kHz
IEC 61000-4-8 Magnetic Field Immunity
30A/m, 50 or 60Hz
IEC 61000-4-11 Voltage Dips
0% U
T
per 0.5 cycles, 0% U
T
per 1.0 cycle, 70% U
T
per 25/30
cycles
IEC 61000-4-11 Voltage Interruptions
0% U
T
per 250/300 cycles