Presence of fibrotic tissue may require as much as 80 mmHg in order
to soften the fibrotic tissue and achieve reduction. Once the tissue is
soft, the compression can be readjusted to 50 mmHg.
Patients with a history of Congestive Heart Failure (CHF), which is
controlled with medication, should never be in a flat position while
pumping. They should be in a reclined position with elevated legs
during treatment. Their treatment regimen duration may be divided into
twice a day 30 minutes per treatment.
Patients with a history of Deep Vein Thrombosis with or without a filter
may require less compression. These patients will generally tolerate 40
mmHg. Patients with a filter may need to divide their treatment into
twice a day, 30 minutes per treatment. It is suggested that the provider
obtain a negative Doppler study from the physician for their records.
Device Description and Operating Principle
Sequential Circulators provide gradient pneumatic compression for the
treatment of lymphedema and associated venous disorders. Sequential
gradient compression helps to increase blood flow and move excess lymph
away from the affected area for clearance from the body. This device can
provide sequential (distal to proximal) inflation/deflation cycles of
compressed air at prescribed pressures. Pneumatic compression devices
are proven to reduce limb swelling associated with lymphedema, help close
up chronic ulcers, and act as a prophylaxis for Deep Vein Thrombosis
(DVT).
Adjustable Pressure
The pressure of the pump can be adjusted between 10 and 120 mmHg.
Pressure can be selected before treatment and adjustments can be made
during treatment.
Adjustable Cycle Time
The cycle time is the time it takes the pump to inflate and deflate the
garment. The cycle time can be adjusted from 60 to 120 seconds in 15
second intervals.